QuickCODlab
LARLAR

QuickCODlab

Bench-top laboratory COD analyser using 1,200 °C thermal oxidation to determine chemical oxygen demand in three minutes, without reagents or catalysts.

laboratory

The QuickCODlab brings the high-temperature method used for on-line COD monitoring into the laboratory. Following ASTM D6238-98 but operating at 1,200 °C instead of the conventional 900 °C, the analyser oxidises all organic compounds without catalysts, hazardous chemicals or corrosive reagents. The sample is injected directly into the reactor and the oxygen consumed is measured with an O2 gas detector.

A complete measurement takes about three minutes, so triplicate results are possible in under ten minutes compared with the hours required by the standard dichromate procedure. A precision syringe with a thumb-wheel dose adjustment and spring-loaded plunger removes operator variation, and no pre-filtration is needed, keeping the sample representative.

Five pre-defined ranges cover 100, 200, 1,000, 3,000 and 10,000 mg/L COD, and user-defined ranges can be stored and recalled for laboratories running samples from many different sources. Salt concentrations up to 10 g/L are handled as standard, and up to 300 g/L sodium chloride with the high-salt furnace, without dilution.

Features and benefits

  • Chemical-free, reagent-free COD analysis
  • Results in three minutes
  • Determination of the real oxygen demand
  • Highest combustion temperature available (1,200 °C)
  • No filtration and no memory effects
  • Excellent chloride acceptance
  • Very low maintenance and operational costs

Applications

Industrial effluent monitoring and influent load monitoring at municipal wastewater treatment works.

Specifications

Measurement techniqueCombustion, ASTM D6238-98 at 1,200 °C
Measuring ranges1–100, 10–200, 100–1,000, 1,000–3,000 and 2,000–10,000 mg/L COD; user-configurable ranges available
Response time3 minutes
Sample injectionManual precision syringe, 1–200 µL dose
Salt toleranceUp to 10 g/L as standard; up to 300 g/L NaCl with high-salt furnace
HousingPowder-coated steel IP 54
Dimensions19.3 × 16.5 × 17.3 in (W × H × D)
Weight81.5 lb
Power supply115 / 230 VAC, 50/60 Hz
Safety2/6 A internal, 16 A external
SoftwareLAR QuickCODlab, Windows 7 or higher
Data interfaceUSB 2.0, LAN (optional)
Data outputExport as .csv file

Options

  • ▪ 60-vessel automatic sampler
  • ▪ High-salt furnace up to 300 g/L NaCl

Enquire about the QuickCODlab

Tell us about your process, sample or monitoring requirement and we will recommend the right configuration. Attach your technical specification or tender document and an application specialist will respond.

  • ▪ Application-matched configuration advice
  • ▪ Datasheets, certificates and drawings
  • ▪ Quotation with lead times